Description
The UT256A digital multimeter is a safe and reliable fork meter that can replace a clamp multimeter. It has no mechanically moving parts and this results in its resistance to mechanical failures. The device has a display with a maximum display value of 6000. Using a fork with a Hall probe allows the measurement of AC current up to 200A. Using the included measuring cables, it is also possible to measure AC and DC voltage, resistance, wire continuity, it has a built-in non-contact AC voltage sensor (NCV) and a built-in lamp. It has data pause functions, automatic and manual range selection, backlit display, frequency measurement, low battery indication. The maximum diameter of the wire that can be inserted into the fork is 15mm. Indication of range exceeding by red LED, for NCV by acoustic signal. The advantage is resistance to falling from a height of up to 2m. Powered by two 1.5 V AA batteries. These features make the UT256A multimeter a practical device for the daily work of high-voltage electricians.
Product parameters :
- measurement of AC current up to 200 A, and its frequency in the range of 50 Hz to 60Hz
- measurement of AC voltage up to 1000 V, and its frequency in the range of 10 Hz to 10 kHz
- DC voltage measurement up to 1000 V
- resistance measurement up to 60 MΩ
- line continuity check
- automatic polarity display
- option to select automatic or manual range switching
- built - in lamp
- True RMS
- the maximum value of the display is 6000
- backlit LCD display measuring 35 mm x 26 mm
- hold
- powered by two 1.5 V AA batteries
- low battery indication
- automatic switch-off after 15 minutes (the function can be temporarily switched off)
Technical parameters :
- AC current measurement :
( ranges / resolution / accuracy)
200.0 A / 0.1 A / ± (1.8 % + 2) in the range of 0.2 A to 2.0 A, 50 to 60 Hz
200.0 A / 0.1 A / ± (1.8 % + 3) in the range of 2,0 A to 5,0 A, 50 to 60 Hz
200.0 A / 0.1 A / ± (1.8 % + 5) in the range of 5,0 A to 200 A, 50 to 60 Hz
- frequency monitoring accuracy when measuring AC current above 20 A, in the range of 50 to 60 Hz is ± (0.1% + 5) at a resolution of 0.01 Hz
- overload protection at 200 A
- Měření AC napětí :
( ranges / resolution / accuracy )
6,000 V / 0,001 V / ± (1,2 % + 3), in the range of 10 Hz to 10 kHz
60,00 V / 0,01 V / ± (1,2 % + 3), in the range of 10 Hz to 10 kHz
600,0 V / 0,1 V / ± (1,2 % + 3), in the range of 10 Hz to 10 kHz
1000,0 V / 1 V / ± (1,2 % + 8), in the range of 10 Hz to 10 kHz
- measures the real rms voltage value in the frequency range 45 Hz to 400 Hz
- frequency monitoring accuracy when measuring AC voltage in the range of 10 Hz to 10 kHz is ± (0.1% + 5) at a resolution of 0.01 Hz to 0.01 kHz, at an amplitude greater than 8 V RMS
- overload protection at 1000 Vrms
- input impedance approx. 10 MΩ
- DC voltage measurement :
( ranges / resolution / accuracy )
6,000 V / 0,001 V / ± (0,8 % +3)
60,00 V / 0,010 V / ± (0,8 % +3)
600,0 V / 0,100 V / ± (0,8 % +3)
1000,0 V / 1 V / ± (1,0 % +5)
- overload protection at 1000 Vrms
- input impedance approx. 10 MΩ
- Resistance measurement :
( ranges / resolution / accuracy )
600,0 Ω / 0,1 Ω / ± (1,2 % + 2)
6,000 kΩ / 0,001 kΩ / ± (1,0 % + 2)
60,00 kΩ / 0,01 kΩ / ± (1,0 % + 2)
600,0 kΩ / 0,1 kΩ / ± (1,0 % + 2)
6,000 MΩ / 0,001 MΩ / ± (1,2 % + 2)
60,00 MΩ / 0,01 MΩ / ± (2,0 % + 5)
- overload protection at 1000 Vrms
- open circuit voltage approx. 1V
- Circuit continuity check :
- if the measured resistance is less than 10 Ω, the signal of the continuous circuit sounds a buzzer
- if the measured resistance is greater than 100 Ω, the open circuit signal does not sound a buzzer
- open circuit voltage about 1.0V
- measurement in the range of 600 Ω, with a resolution of 0.1Ω
- overload protection at 1000 Vrms
- Non-contact electric field sensing (NCV) :
- low voltage detection mode ("EFLo") for voltage <48V
- low voltage detection mode ("EFHi") for voltages >48V
- as the intensity of the detected voltage increases, the number of segments on the display increases, the green, yellow and red LEDs flash successively and a buzzer sounds
General parameters :
- working temperature and ambient humidity: 0 °C ~ 30 °C (≤80% RH), 30 °C ~ 40 °C (≤75% RH), 40 °C ~ 50 °C (≤45% RH)
- storage temperature and humidity: -20 °C ~ 60 °C (≤80% RH)
- working altitude: ≤2000m
- electromagnetic compatibility: in the elmag. field RF = 1V / m there is an additional error of 5% of the range, in the field RF >1V / m, the total accuracy is not specified
- product dimensions: 217 x 58.8 x 38.5 mm
- net weight of the product: 251 g
- product color: red and gray
- this device is designed in accordance with safety standards IEC / EN61010-1, IEC / EN61010-2-033 and EN61326-1 and complies with CAT II 1000V, CAT III 600V, double insulation.
Package contents :
- clamp multimeter UT256A: 1pc
- test leads: 1 pair
- battery 1.5 V (LR6): 2pcs
- instructions for use: 1pc
Wholesale packaging :
- standard quantity per carton: 20pcs
- standard carton dimensions: 485 x 310 x 205 mm
- standard gross carton weight: 9.5 kg
MANUFACTURER: UNI-TREND
https://www.uni-trend.com/
Documents available for download
Nobody has entered comment on this product yet. Be first!